PerfectLum vs Verilum: a comparison

of medical monitor calibration tools


Features QUBYX PerfectLum 3.x Image-Smith Verilum 5.2
Remote display management system with web interface Yes No
Remote QA Yes with PerfectLum 3.x Suite No
Scheduler Yes No
Type of connected display device Detected automatically Needs to be manually set
Device settings like clone, mirror, single Automatic Manual
Automatic Display Adjustment over DDC/CI Yes in DDC/CI compatible displays No
DICOM calibration Yes Yes
CIE L* calibration Yes No
Gamma calibration Yes Yes
VESA DDC/CI communication Yes No
Multilingual Yes No
Automatic sensor installation Yes No
Sensor support Colorimeter
  • X-Rite Eye-One display
  • X-Rite Eye-One display 2
  • X-Rite Monaco OPTIX XR
  • X-Rite DTP 94
  • X-Rite Gamma 3
  • X-Rite Chroma 4
  • X-Rite Chroma 5
  • Datacolor Spyder3
Spectralphotometer
  • Gretag Macbeth Eye-One Pro
  • Gretag Macbeth Eye-One Monitor
Spot Luminance meter
  • IBA dositometry / Wellhoefer LX Plus
  • IBA dositometry / Wellhoefer LX Can
  • IBA dositometry / Wellhoefer LX Chroma
  • Pehamed CD mon
  • Pehamed CD LUX
  • Pehamed CD Lux plus

Display Front Sensors

  • LumiMED MM03A
  • LumiMED MM05A
  • Olorin VC19012
  • Olorin MC20012
  • Olorin VC21012
Verilum photometer
Number of patch selection for calibration Dynamic Manual
Screen size settings Automatically adjusts to the desktop screen size Set in preferences
DeltaE verification Yes No
Measure white level Yes Yes
Measure black level Yes Yes
Measure luminance uniformity Yes Yes
Analyse Pass/Fail of uniformity for AAPM TG18, DIN6565-57, JESRAX standard Yes No
Measure luminance Yes Yes
Measure color Yes No
Calculate Delta E between two colors and gray levels Yes No
Display test patterns
  • Briggs
  • SMPTE
  • AAPM TG-18
  • TG18-QC
  • TG18-AD
  • TG18-CH
  • TG18-CT
  • TG18-KN
  • TG18-MP
  • TG18-PDS
  • TG18-MM01
  • TG18-MM02
  • IEC DIN 6868-57
  • 8bit Grayramp
  • Gray scale
  • Gretag Color checker preinstalled and drawn for any resolution
SMPTE but does not fit screen for any resolution, "Verilum" pattern
User defined patterns Yes Yes
Resultant graphs
  • Luminance response graph.
  • Gamma correction curve.
  • JND conformance.
  • AAPM conformance.
  • DeltaE graph.
  • Gamma deviation
  • Uniformity graph
  • Color temperature consistency graph
  • Luminance response
  • Gamma correction
  • JND conformance
Matrox med and RAD series Supported Supported
Matrox XENIA and XENIA Pro board Supported Not supported
Uniformity adjustment with Matrox XENIA board Yes No
Reference calibration Yes No
Detailed history Yes No. Only summary. User can only see details of the last calibration or conformance check
Color calibration over gray scale Yes No
DICOM calibration Yes Yes
Generates ICC profiles Yes No
Measures luminance for Gamma correction Yes Yes
Consistency graphs for white and black levels Yes No
DICOM conformance Yes Yes
Different color temperature settings can be set. Yes No
Preference setting Less settings and simple Manual settings needed
Platform Windows 2000, XP, Vista, Windows 7 32 and 64 bit, Mac OSX Windows 2000, XP, Vista, Windows 7 (for single display only)
AAPM TG18/DIN 6868-57 acceptance and conformance Yes with PerfectLum 3.x Suite No
IEC 62563-1/JESRA X-0093 acceptance and conformance
Yes with PerfectLum 3.x Suite No